TL1200-1200 double temperature zone fast heating furnace

  • Detailed

 1. Kanthal A1 resistance wire of Sweden

             

           

        Kanthal A1               Chinese                          Imported

(1), high surface temperature can reach 1420 ℃

(2) the surface is stainless steel (very bright, the customer should pay attention to it), it will not rust, and the use time will not fall off the slag.

(3) data of Kanthal resistance wire is downloaded

(4) Kanthal resistance wire website: www.kanthal.com

(5) the resistance is balanced and the temperature field is equally good

                                                                                                                  

Most of the manufacturers use fe-cr-al of

shougang production Alloy doped by Mo                       Kanthal  A1 Resistance wire is uniform                   Resistance wire sometimes has an uneven resistance and

a portion of the heat is less heat                                   Good heat balance temperature                                                                                    

 

2. Installation of air switch and leakage protector: When an electric current or leakage, an empty meeting automatically disconnects.

 

3. Can connect with computer

This furnace is equipped with communication interface and software, can be directly through the furnace computer control of each parameter, and can be observed from the computer PV on the stove and SV temperature value and instrument operation, the actual temperature of the furnace curve drawing computer real time, and can keep the temperature data of each time, can call up at any time

 

 

4. Chamber

 

     

1. High quality high pure alumina polycrystalline fiber cured furnace with vacuum suction and filter molding.

2. Adopt Japanese technology to form.

3. The spacing and pitch of the resistance wire in the furnace are arranged according to the best thermal technology of Japan, and the temperature field is simulated by thermal software

4. Using 4 weeks heating, the temperature is more balanced

 

5. Optional UL certified electrical board (all imported electrical appliances approved by UL)

 

 

6.Certification

     

 

7、Bottum studs

         

8、Alumina furnace plug

         

Outside the tube furnace with resistance wire Canthal A1 (Sweden) as the heating element, using double shell structure and 30 program temperature control system, the phase shift trigger, thyristor control, furnace adopts polycrystalline alumina fiber materials, low temperature field balancing, surface temperature and elevating temperature rate, energy saving, colleges and universities, research institutes, industrial and mining enterprises to do the high temperature sintering atmosphere, reducing atmosphere, CVD, vacuum annealing test with the ideal product.

 

Detailed technical parameters

Model

TL1200-1200-Q-I

TL1200-1200-Q-II

TL1200-1200-Q-III

TL1200-1200-Q-IV

Power

2.5KW

2.5KW

 4KW

4KW

pipe size(OD(mm)

Dia 60(OD) X1460mm

Dia80(ODX1460mm

Dia100(OD)X1460mm

Dia120(OD)X1460mm

boundary dimension 1500X490X600 1500X490X600 1500X490X600 1500X490X600
  • Supply voltage

220V

Phase

Single

Heating element

Swedish Kanthal A1 imported resistance wire (resistance wire guaranteed for 2 years)   

  • Temperature control type

Using the yu electric program control thermometer (standard)

1. 30 program control temperature intelligent PID adjustment.

2, over temperature protection is broken accidentally protection with a temperature or broken accidentally electric furnace heating circuit automatically cut OFF, and, when the temperature of the furnace more than 1200 degrees or thermocouple burn out, the main circuit of ac relay will automatically disconnect, main circuit disconnect, panel ON lights, OFF the light, the limited protection electric furnace).

3. With 485 communication interface (standard and standard for choosing software)

4. It has the function of power failure protection, that is, when the power is off, the program will not start up from the beginning temperature, but the furnace will start to rise when the power is off.

5. The instrument has the function of temperature self-tuning

(Optional)

                                                                                                

FP93 SHIMADEN,Japan                            EUROTHERM                    Touch screen instrument

Chamber materials

1. High quality high pure alumina polycrystalline fiber cured furnace with vacuum suction and filter molding.

2. Adopt Japanese technology to form.

3. The spacing and pitch of the resistance wire in the furnace are arranged according to the best thermal technology of Japan, and the temperature field is simulated by thermal software

  • Flange

Control precision

 +/- 1 ℃

  • Trigger

 Phase-shifted trigger

Electric appliance

 Zhejiang chint

silicon controlled

106/16E SEMIKRON

maximum temperature

 1200 ℃

 rated temperature

 1100 ℃

Heating length

 205mm/ 205mm

Length of constant temperature

80mm-100mm

Heating rate

≤100 ℃/ Min (adjustable)

  • Thermocouple Type

 K

Surrounding surface temperature

≤45℃

Open cover protection system When the cover is opened, the relay will automatically cut off the main power supply to ensure the user's safety.

Vacuum system (Optional)

                                                            

  1、single stage pumpTW-1.5A     2、double-stage pump 2XZ-2     3、 VS-0.1vacuum system  

 

                    

     4.VS high vacuum system        5、 constant pressure pump      

  

  6、English PCG550(INFICON)Capacitance manometer

1. Generally, the gas in the tube should be driven out of the tube in A general way: the tw-1.5 A single-stage pump can be used.

2. Two-stage pump 2xz-2 vacuum pump can pump the vacuum of tube furnace tube to about 10Pa.

3. The vs-0.1 vacuum system can pump out the vacuum in tube furnace tube to 0.1 Pa (with resistance vacuum gauge display).

4. VS high vacuum system (which is composed of molecules and mechanical pumps) can vacuum the tube furnace tube to 10-3 Pa.

5. Constant pressure pump (the system can control the internal pressure of the tube to -100.0 KPa ~ 100.0 KPa).

6. British (INFICON) capacitor vacuum gauge (measurement range of 3.8 x10-5-1125 Torr, which can detect various gases, including mixing gas, accurate, repeatable, optional anti-corrosion type PCG554)

Gas supply system (Optional)

                    

           1、 bubbler        2、Mass flowmeter supply system        3、Liquid evaporator

 



Tel/whatsapp/wechat: +86 13913006686